INTRODUCTION
CONFIGURATIONS OF FUME HOOD & EXTRACTION SYSTEM

FUME HOOD WITH WET SCRUBBER SYSTEM

FUME HOOD WITH DRY FILTERS SYSTEM

Pre-filter plus Activated Charcoal Filter, or 
Pre-filter plus H.E.P.A.Filter
 

 

TECHNICAL PLANNING OF FUME HOODS

LOW EFFICIENCY LOCATION

 

PREFERRED LOCATION



GAS DETECTION SYSTEM FOR FUME HOODS
A system consists of the components necessary for continuous monitoring of gaseous substances for 24 hours a day can be incorporated into 'MODULAP' Fume Cupboard System. 
  • The primary purpose for having a Gas Detection System is to insure early 'WARNING' of the presence of either a combustible or a toxic gas. Once the presence is known, action can be taken to disperse the gas before it ignites or overcomes occupants.
Contact our representative our technical advice and recommendation.

 

 

'MODULAP' fume hood range which was established since 1983 has been proven to operate more efficiently than similar unit of a more traditional design. It's aerodynamically designed streamlined angled entrance allows air to enter in smooth unbroken pattern, permitting operation with greater safety and reduce risk of frontal turbulence, of which it may result in a higher containment of the fumes.

 

GENERAL FEATURES

 

One piece moulded chamber eliminates joints and provides for easy wash down and decontamination. Liner with radiused corners to maintain a turbulence free air flow pattern.
The white colour moulded chamber also assists in evenly illuminating the working area and gives the operator a pleasant working environment.


Access to service fixture piping is concealed between the Inner and Outer walls, and the removable Access Panels on both sides of the hood interior make it possible to install additional service fixture for the fume hood with least obstruction for future needs.


The fluorescent lighting fixture is mounted inside the moulded FRP casing at exterior of the fume hood roof, and keeps away from the exhaust air stream. The safety glass shield is provided with air-tight silicon sealant to isolate the light fitting from the hood interior.





All  the Electrical services are mounted on the front Service Panels above the work-top level. Hence the chances of spillage by liquid/regent on the working top not any electrical Point which may cause Electrical Short-circuited is virtually impossible. Furthermore, all the electric PVC insulated cable-circuitry for the fume hood is further protected by additional good quality & chemical-resistant plastic conduit.


A chemical resistant Airfoil Panel is provided on top-front facial of the fume hood  insuring a constant sweep across the work surface of efficient fume removal.




An automatic air by-pass grille is provided on top-front facial of the fume hood to ensure that the face velocity is not extensive at small sash opening, with fumes are adequately diluted, and the exhaust air volume from the laboratory is relatively kept constant.


 
The safety glass sash window which is balanced by concealed counterweights on Vinyl coated stainless steel ropes and guided by nylon ball bearing pulleys can be sliding vertically with least effort of control at any sash opening.
TECHNICAL INFORMATION
Laboratory fume hoods shall provide safe operation when properly installed and connected to an exhaust system that provides the proper exhaust air volume to permit the fume hood to operate at the specified face velocity.

'MODULAP' fume cupboards when utilized with the recommended extraction system conforms to the industrial standards for safe velocities. The table shown below summarizes the recommended face velocities for different laboratory conditions.

FUMES/LABORATORY RELATIVE HAZARD OR TOXICITY AVE. FACE VELOCITY AT THE FACE OF THE FUME HOODS
Heat, steam, smoke odour, etc Low 75-80      (Foot-per-min)
General Chemical Laboratory Intermediate 100          (Foot-per-min)   
Concentrated & strong chemicals Of extreme toxicity or hazardous. High 125-150  (Foot-per-min)
Fume Hoods used Isotope materials, Explosive Chemicals/Reagents. High 125-150  (Foot-per-min)
DUCTWORK VARIATIONS
AERO STREAM BY-PASS BENCH TOP GENERAL PURPOSE FUME HOOD (FCS MODEL)
Superstructure are available in size ranging from four to eight foot wide designed for installation on a 30" deep counter top.

Size (wide)
Fume Hood Cat No.
Dimension W x D x H
1220mm
FCS-1200-BP
1220x914x1525mm
1525mm
FCS-1500-BP
1525x914x1525mm
1830mm
FCS-1800-BP 1830x914x1525mm
2440mm
FCS-2400-BP
2440x914x1525mm
Base Unit Cat No.
Dimension W x D x H
(Free Standing)
SS-4834-F
1220x762x860mm
SS - 6034-F
1525x762x860mm
SS - 7234 - F
1830x762x860mm
SS - 9634 -F
2440x762x860mm
Exhaust Dust size 250mm. Dia. 250mm. Dia. 300mm. Dia. two 250mm. Dia.
Usable Chamber Size
Dimension W x D x H
1016x660x1143mm 1320x660x1143mm 1625x660x1143mm 2235x660x1143mm
Sash Opening Size
Dimension W x H
1016x700mm 1320x700mm 1625x700mm 2235x700mm
'MODULAP'  Fume Hood based on 100 fpm (30m/min) face velocity with duct static pressure less than the equivalent of 50ft (15.2m) of straight duct.

(OTHER DIMENSIONS ARE POSSIBLE ON REQUEST)

 

 

AERO STREAM WALK-IN FUME HOOD FCW MODEL
Walk-In floor standing hood is equipped where height is a main consideration with Dual Sash to allow 1,950mm clearance height. Otherwise it is similar to FCS models.

Size (wide)
Fume Hood Cat No.
Dimension W x D x H
1220mm
FCS-1200-BP
1220x914x2385mm
1525mm
FCS-1500-BP
1525x914x2385mm
1830mm
FCS-1800-BP 1830x914x2385mm
2440mm
FCS-2400-BP
2440x914x2385mm
Exhaust Dust size 300mm. Dia. Two 250mm. Dia. Two 300mm. Dia. Three 250mm. Dia.
Usable Chamber Size
Dimension W x D x H
1016x660x2055mm 1320x660x2055mm 1625x660x2055mm 2235x660x2055mm
Sash Opening Size
Dimension W x H
1016x1650mm 1320x1650mm 1625x1650mm 2235x1650mm
AERO STREAM DISTILLATION FUME HOODS FCD MODEL
Distillation hood has dual sash opening allowing up to 1,500mm clearance height. 450mm height of low Base Cabinet is incorporated with this model. Otherwise it is similar to FCS models.

Size (wide)
Fume Hood Cat No.
Dimension W x D x H
1220mm
FCS-1200-BP
1220x914x1928mm
1530mm
FCS-1500-BP
1530x914x1928mm
1830mm
FCS-1800-BP 1830x914x1928mm
2440mm
FCS-2400-BP
2440x914x1928mm
Base Unit Cat No.
Dimension W x D x H
(Free Standing)
SS-4834-F
1220x760x457mm
SS - 6034-F
1530x760x457mm
SS - 7234 - F
1830x760x457mm
SS - 9634 -F
2440x760x457mm
Exhaust Dust size 300mm. Dia. Two 250mm. Dia. Two 300mm. Dia. Three 250mm. Dia.
Usable Chamber Size
Dimension W x D x H
1016x660x1600mm 1320x660x1600mm 1625x660x1600mm 2235x660x1600mm
Sash Opening Size
Dimension W x H
1016x1194mm 1320x1194mm 1625x1194mm 2235x1194mm
BASE CABINETS FOR BENCH TOP MODEL
Free Standing Base Cabinet :


 

As shown above.
The base cabinets are made of Epoxy powder coated Steel with PVC sheet lined on internal shelving and designed with vent hole.
CAT. NO.
SS - 4834- *S/W
SS - 6034- *S/W
SS - 7234- *S/W
SS - 9634- *S/W
DIMENSION: WxDxH
1,200 x 760 x 850
1,520 x 760 x 850
1,820 x 760 x 850
2,440 x 760 x 850
Suspended Base Cabinets  :



 

 

 

('*S' : Epoxy powder coated, or '*W':Wooden)
Snap-on suspended base cabinets are designed to hang beneath cantilevered, and the epoxy powder coated steel table frame-work and are furnished with all hardwares required for attachment to the metal frames. The base cabinet can be replaced easily without dismantling of fume hood and duct works. The internal shelving of all base cabinet are lined with vent hole.

OTHER CONFIGURATION
1. AUXILIARY-AIR FCS MODEL FUME HOOD

'MODULAP' General purpose Fume Hood with a distribution box added on top of the fume hood can convert it into an Auxiliary Air Models.

This model is often used in air-conditioned laboratories. It can save both initial outlay and continued operating costs. The Auxiliary Air mo Model is normally set to take 50% of air from the laboratory. As a result, it provides an even greater degree of efficiency and cost saving in operation. Generally, architects estimated that every 200 cubic feet per minute of cooled air exhausted through a fume hood requires an extra tons of air-conditioning capacity.

For example, the standard 4-ft By-Pass model operating at a face velocity of 100 fpm loses approximately 760 cubic feet of cooled air per minute, thus will need 3.8 tons of extra air-conditioning.

The Auxiliary Air model, on the other hand, will require only 50% as much extra air-conditioning capacity.

     The result : A saving of up to 50% on air conditioning expenses.

A separate motor-blower to draw the outside air into a Distribution Box on the top of the Fume Hood is needed for this Auxiliary Air Fume Hood.

 

2. VARIABLE AIR VOLUME (VAV) CONTROL SYSTEM - THE ENERGY SAVER

Variable Speed Control circuitry served to vary or set to the required face velocity of the fume cupboard exhaust system is an Optional feature to the fume cupboard system. The speed Controller's terminal is attached to the Sash Mechanism system so as to achieve energy cost savings due to reduction in extracted conditioned air in the laboratory. The amount of exhaust air varies in proportion to the sash opening, while maintaining almost constant face velocity regardless of sash position. This configuration is generally not affected by temperature or humidity changes. (This is possible only when the by-pass system has to be concealed completely).

Our low cost VAV Control System has proof popular in general applications with adjustable face velocity flexibility.

3. EXPLOSION - PROOF FUME HOOD
'MODULAP' GENERAL PURPOSE Fume Hood can be converted into Explosion - proof model consists of Stainless Steel Inner Liner, Explosion - proof Lightings, Fittings and Electrical Sockets. The electric cables are made of Fire Resistant Cables in accordance with IEE Reg. 423-1 and 523-5 (Meet IEC 331 Flame Test).
4. SPECIAL APPLICATIONS FUME HOODS
  • Perchloric Acid Fume Hoods
  • Hydrofluoric Acid Fume Hoods
  • Class 1 Safety Cabinets
ALSO AVAILABLE - ON SPECIAL REQUEST:
  • Dual sash Postless Fume Hoods.
  • Mobile Fume Cabinet for school.
  • Horizontal sliding panels - added to vertical frame sash.
  • Glove Port Rings sash.
  • Custom built Models for Special Applications. (With Fume Hoods using Stainless Steel, Polypropylene, or PVC materials as Inner Liner & Baffle.

Contact our representative for technical advice and recommendation. 

 

 